Management of patients with cardiac implantable electronic devices (CIED) undergoing radiotherapy

A consensus document from Associazione Italiana Aritmologia e Cardiostimolazione (AIAC), Associazione Italiana Radioterapia Oncologica (AIRO), Associazione Italiana Fisica Medica (AIFM)
Published:December 22, 2017DOI:https://doi.org/10.1016/j.ijcard.2017.12.061

      Abstract

      The management of patients with a cardiac implanted electronic device (CIED) receiving radiotherapy (RT) is challenging and requires a structured multidisciplinary approach. A consensus document is presented as a result of a multidisciplinary working group involving cardiac electrophysiologists, radiation oncologists and physicists in order to stratify the risk of patients with CIED requiring RT and approaching RT sessions appropriately.
      When high radiation doses and beam energy higher than 6 MV are used, CIED malfunctions can occur during treatment. In our document, we reviewed the different types of RT and CIED behavior in the presence of ionizing radiations and electromagnetic interferences, from the cardiologist's, radiation oncologist's and medical physicist's point of view. We also reviewed in vitro and in vivo literature data and other national published guidelines on this issue so far. On the basis of literature data and consensus of experts, a detailed approach based on risk stratification and appropriate management of RT patients with CIEDs is suggested, with important implications for clinical practice.

      Keywords

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to International Journal of Cardiology
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Harms W.
        • Budach W.
        • Dunst J.
        • et al.
        DEGRO practical guidelines for radiotherapy of breast cancer VI: therapy of locoregional breast cancer recurrences.
        Strahlenther. Onkol. 2016; 192: 199-208
        • Soejima T.
        • Yoden E.
        • NI Y.
        • et al.
        Radiation therapy in patients with implanted cardiac pacemakers and implantable cardioverter defibrillators: a prospective survey in Japan.
        J. Radiat. Res. 2011; 52: 516-521
        • Zaremba T.
        • Jakobsen A.R.
        • Sogaard M.
        • et al.
        Risk of device malfunction in cancer patients with implantable cardiac device undergoing radiotherapy: a population-based cohort study.
        Pacing Clin. Electrophysiol. 2015; 38: 343-356
        • Hurkmans C.W.
        • Knegjens J.L.
        • Oei B.S.
        • et al.
        Management of radiation oncology patients with a pacemaker or ICD: a new comprehensive practical guideline in The Netherlands. Dutch Society of Radiotherapy and Oncology (NVRO).
        Radiat. Oncol. 2012; 7: 198
        • Indik J.H.
        • Gimbel J.R.
        • Abe H.
        • et al.
        2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices.
        Heart Rhythm. 2017; 14: e97-e153
        • Hindricks G.
        • Camm J.
        • Merkely B.
        • Raatikainen P.
        • Arnar D.O.
        The EHRA White Book 2016 — The Current Status of Cardiac Electrophysiology in ESC Member Countries.
        2016
        • Marbach J.R.
        • Sontag M.R.
        • Van Dyk J.
        • Wolbarst A.B.
        Management of radiation oncology patients with implanted cardiac pacemakers: report of AAPM Task Group No. 34. American Association of Physicists in Medicine.
        Med. Phys. 1994; 21: 85-90
        • Sundar S.
        • Symonds R.P.
        • Deehan C.
        Radiotherapy to patients with artificial cardiac pacemakers.
        Cancer Treat. Rev. 2005; 31: 474-486
        • Wadasadawala T.
        • Pandey A.
        • Agarwal J.P.
        • et al.
        Radiation therapy with implanted cardiac pacemaker devices: a clinical and dosimetric analysis of patients and proposed precautions.
        Clin Oncol (R Coll Radiol). 2011; 23: 79-85
        • Hudson F.
        • Coulshed D.
        • D'Souza E.
        • Baker C.
        Effect of radiation therapy on the latest generation of pacemakers and implantable cardioverter defibrillators: a systematic review.
        J Med Imaging Radiat Oncol. 2010; 54: 53-61
        • Tondato F.
        • Ng D.W.
        • Srivathsan K.
        • Altemose G.T.
        • Halyard M.Y.
        • Scott L.R.
        Radiotherapy-induced pacemaker and implantable cardioverter defibrillator malfunction.
        Expert Rev Med Devices. 2009; 6: 243-249
        • Langer M.
        • Orlandi E.
        • Carrara M.
        • Previtali P.
        • Haeusler E.A.
        Management of patients with implantable cardioverter defibrillator needing radiation therapy for cancer.
        Br. J. Anaesth. 2012; 108 (author reply 2): 881-882
        • Gauter-Fleckenstein B.
        • Israel C.W.
        • Dorenkamp M.
        • et al.
        DEGRO/DGK guideline for radiotherapy in patients with cardiac implantable electronic devices.
        Strahlenther. Onkol. 2015; 191: 393-404
        • Cardio-Oncology Iwgo
        Cardio-oncologia.
        2015
        • Stone M.E.
        • Salter B.
        • Fischer A.
        Perioperative management of patients with cardiac implantable electronic devices.
        Br. J. Anaesth. 2011; 107: i16-26
        • Giaj-Levra N.
        • Sciascia S.
        • Fiorentino A.
        • et al.
        Radiotherapy in patients with connective tissue diseases.
        Lancet Oncol. 2016; 17: e109-17
        • Lelakowski J.
        • Majewski J.
        • Bednarek J.
        • Malecka B.
        • Zabek A.
        Pacemaker dependency after pacemaker implantation.
        Cardiol J. 2007; 14: 83-86
        • Nagatomo T.
        • Abe H.
        • Kikuchi K.
        • Nakashima Y.
        New onset of pacemaker dependency after permanent pacemaker implantation.
        Pacing Clin. Electrophysiol. 2004; 27: 475-479
        • Rosenheck S.
        • Bondy C.
        • Weiss A.T.
        • Gotsman M.S.
        Comparison between patients with and without reliable ventricular escape rhythm in the presence of long standing complete atrioventricular block.
        Pacing Clin. Electrophysiol. 1993; 16: 272-276
        • Hussein A.A.
        • Abutaleb A.
        • Jeudy J.
        • et al.
        Safety of computed tomography in patients with cardiac rhythm management devices: assessment of the U.S. Food and Drug Administration advisory in clinical practice.
        J. Am. Coll. Cardiol. 2014; 63: 1769-1775
        • Bourgouin A.
        • Varfalvy N.
        • Archambault L.
        Estimating and reducing dose received by cardiac devices for patients undergoing radiotherapy.
        J Appl Clin Med Phys. 2015; 16: 5317
        • Munshi A.
        • Agarwal J.P.
        • Pandey K.C.
        Cancer patients with cardiac pacemakers needing radiation treatment: a systematic review.
        J. Cancer Res. Ther. 2013; 9: 193-198
        • Isobe T.
        • Kumada H.
        • Takada K.
        • et al.
        Effects of secondary neutron beam generated in radiotherapy on electronic medical devices.
        Progress in Nuclear Sci. Technol. 2011; 2: 524-529
        • Klein E.E.
        • Maserang B.
        • Wood R.
        • Mansur D.
        Peripheral doses from pediatric IMRT.
        Med. Phys. 2006; 33: 2525-2531
        • Kry S.F.
        • Bednarz B.
        • Howell R.M.
        • et al.
        AAPM TG 158: measurement and calculation of doses outside the treated volume from external-beam radiation therapy.
        Med. Phys. 2017; 18: 715-716
        • Tajstra M.
        • Gadula-Gacek E.
        • Buchta P.
        • Blamek S.
        • Gasior M.
        • Kosiuk J.
        Effect of therapeutic ionizing radiation on implantable electronic devices: systematic review and practical guidance.
        J. Cardiovasc. Electrophysiol. 2016; 27: 1247-1251
        • Schneider U.
        • Halg R.
        The impact of neutrons in clinical proton therapy.
        Front. Oncol. 2015; 5: 235
        • Ueyama T.
        • Arimura T.
        • Ogino T.
        • et al.
        Pacemaker malfunction associated with proton beam therapy: a report of two cases and review of literature—does field-to-generator distance matter?.
        Oxf Med Case Reports. 2016; 8
        • Hauser R.G.
        • Kallinen L.
        Deaths associated with implantable cardioverter defibrillator failure and deactivation reported in the United States Food and Drug Administration manufacturer and user facility device experience database.
        Heart Rhythm. 2004; 1: 399-405
        • Brambatti M.
        • Mathew R.
        • Strang B.
        • et al.
        Management of patients with implantable cardioverter-defibrillators and pacemakers who require radiation therapy.
        Heart Rhythm. 2015; 12: 2148-2154
        • Raitt M.H.
        • Bardy G.H.
        Advances in implantable cardioverter-defibrillator therapy.
        Curr. Opin. Cardiol. 1994; 9: 23-29
        • Tsekos A.
        • Momm F.
        • Brunner M.
        • Guttenberger R.
        The cardiac pacemaker patient — might the pacer be directly irradiated?.
        Acta Oncol. 2000; 39: 881-883
        • Nibhanupudy J.R.
        • de Jesus M.A.
        • Fujita M.
        • Goldson A.L.
        Radiation dose monitoring in a breast cancer patient with a pacemaker: a case report.
        J. Natl. Med. Assoc. 2001; 93: 278-281
        • Hoecht S.
        • Rosenthal P.
        • Sancar D.
        • Behrens S.
        • Hinkelbein W.
        • Hoeller U.
        Implantable cardiac defibrillators may be damaged by radiation therapy.
        J. Clin. Oncol. 2002; 20: 2212-2213
        • Frantz S.
        • Wagner J.
        • Langenfeld H.
        Radiation-induced pacemaker malfunction.
        Z. Kardiol. 2003; 92: 415-417
        • John J.
        • Kaye G.C.
        Shock coil failure secondary to external irradiation in a patient with implantable cardioverter defibrillator.
        Pacing Clin. Electrophysiol. 2004; 27: 690-691
        • Thomas D.
        • Becker R.
        • Katus H.A.
        • Schoels W.
        • Karle C.A.
        Radiation therapy-induced electrical reset of an implantable cardioverter defibrillator device located outside the irradiation field.
        J. Electrocardiol. 2004; 37: 73-74
        • Ampil F.L.
        • Caldito G.
        Radiotherapy for palliation of lung cancer in patients with compromised hearts.
        J. Palliat. Med. 2006; 9: 241-242
        • Mitra D.
        • Ghosh K.
        • Gupta P.
        • Jayanti J.
        • Dev A.
        • Sur P.
        Radiation dose monitoring in a lung cancer patient with a pacemaker — a case report.
        Chest. 2006; 16: 875-877
        • Sepe S.
        • Schaffer P.
        • Krimmel K.
        • Schaffer M.
        Irradiation treatment of laryngeal cancer in a patient with an implantable cardioverter-defibrillator (ICD).
        Onkologie. 2007; 30: 378-380
        • Nemec J.
        Runaway implantable defibrillator — a rare complication of radiation therapy.
        Pacing Clin. Electrophysiol. 2007; 30: 716-718
        • Munshi A.
        • Wadasadawala T.
        • Sharma P.K.
        • et al.
        Radiation therapy planning of a breast cancer patient with in situ pacemaker — challenges and lessons.
        Acta Oncol. 2008; 47: 255-260
        • Kapa S.
        • Fong L.
        • Blackwell C.R.
        • Herman M.G.
        • Schomberg P.J.
        • Hayes D.L.
        Effects of scatter radiation on ICD and CRT function.
        Pacing Clin. Electrophysiol. 2008; 31: 727-732
        • Oshiro Y.
        • Sugahara S.
        • Noma M.
        • et al.
        Proton beam therapy interference with implanted cardiac pacemakers.
        Int. J. Radiat. Oncol. Biol. Phys. 2008; 72: 723-727
        • Lau D.H.
        • Wilson L.
        • Stiles M.K.
        • et al.
        Defibrillator reset by radiotherapy.
        Int. J. Cardiol. 2008; 130: e37-8
        • Zweng A.
        • Schuster R.
        • Hawlicek R.
        • Weber H.S.
        Life-threatening pacemaker dysfunction associated with therapeutic radiation: a case report.
        Angiology. 2009; 60: 509-512
        • Gelblum D.Y.
        • Amols H.
        Implanted cardiac defibrillator care in radiation oncology patient population.
        Int. J. Radiat. Oncol. Biol. Phys. 2009; 73: 1525-1531
        • Zaremba T.
        • Thogersen A.M.
        • Eschen O.
        • Hjortshoj S.P.
        • Jakobsen A.R.
        • Riahi S.
        High-dose radiotherapy exposure to cardiac pacemakers may be safe in selected patients.
        Radiother. Oncol. 2010; 95: 133-134
        • Ferrara T.
        • Baiotto B.
        • Malinverni G.
        • et al.
        Irradiation of pacemakers and cardio-defibrillators in patients submitted to radiotherapy: a clinical experience.
        Tumori. 2010; 96: 76-83
        • Dasgupta T.
        • Barani I.J.
        • Roach III, M.
        Successful radiation treatment of anaplastic thyroid carcinoma metastatic to the right cardiac atrium and ventricle in a pacemaker-dependent patient.
        Radiat. Oncol. 2011; 6: 16
        • Menard J.
        • Campana F.
        • Kirov K.M.
        • et al.
        Radiotherapy for breast cancer and pacemaker.
        Cancer Radiother. 2011; 15: 197-201
        • Croshaw R.
        • Kim Y.
        • Lappinen E.
        • Julian T.
        • Trombetta M.
        Avoiding mastectomy: accelerated partial breast irradiation for breast cancer patients with pacemakers or defibrillators.
        Ann. Surg. Oncol. 2011; 18: 3500-3505
        • Kirova Y.M.
        • Menard J.
        • Chargari C.
        • Mazal A.
        • Kirov K.
        Case study thoracic radiotherapy in an elderly patient with pacemaker: the issue of pacing leads.
        Med. Dosim. 2012; 37: 192-194
        • Kesek M.
        • Nyholm T.
        • Asklund T.
        Radiotherapy and pacemaker: 80 Gy to target close to the device may be feasible.
        Europace. 2012; 14: 1595
        • Makkar A.
        • Prisciandaro J.
        • Agarwal S.
        • et al.
        Effect of radiation therapy on permanent pacemaker and implantable cardioverter-defibrillator function.
        Heart Rhythm. 2012; 9: 1964-1968
        • Elders J.
        • Kunze-Busch M.
        • Jan Smeenk R.
        • Smeets J.L.
        High incidence of implantable cardioverter defibrillator malfunctions during radiation therapy: neutrons as a probable cause of soft errors.
        Europace. 2013; 15: 60-65
        • Keshtgar M.R.
        • Eaton D.J.
        • Reynolds C.
        • et al.
        Pacemaker and radiotherapy in breast cancer: is targeted intraoperative radiotherapy the answer in this setting?.
        Radiat. Oncol. 2012; 7: 128
        • Gomez D.R.
        • Poenisch F.
        • Pinnix C.C.
        • et al.
        Malfunctions of implantable cardiac devices in patients receiving proton beam therapy: incidence and predictors.
        Int. J. Radiat. Oncol. Biol. Phys. 2013; 87: 570-575
        • Dell'oca I.
        • Tsiachris D.
        • Oppizzi M.
        • Della Bella P.
        • Gulletta S.
        Radiotherapy and implanted cardioverter defibrillators: novel techniques make it feasible.
        J Cardiovasc Med (Hagerstown). 2007; 34: 1193-1197
        • Zaremba T.
        • Jakobsen A.R.
        • Thogersen A.M.
        • Riahi S.
        • Kjaergaard B.
        Effects of high-dose radiotherapy on implantable cardioverter defibrillators: an in vivo porcine study.
        Pacing Clin. Electrophysiol. 2013; 36: 1558-1563
        • Ampil F.
        • Nathan C.
        • Ghali G.
        • Kim D.
        Postoperative radiotherapy for advanced head and neck cancer in patients with cardiac pacemakers.
        J. Radiother. Pract. 2014; 13: 115-118
        • Gossman M.S.
        • Wilkinson J.D.
        • Mallick A.
        Treatment approach, delivery, and follow-up evaluation for cardiac rhythm disease management patients receiving radiation therapy: retrospective physician surveys including chart reviews at numerous centers.
        Med. Dosim. 2014; 39: 320-324
        • Ahmed I.
        • Zou W.
        • Jabbour S.K.
        High dose radiotherapy to automated implantable cardioverter-defibrillator: a case report and review of the literature.
        Case Rep Oncol Med. 2014; 2014: 989857
        • Grant J.D.
        • Jensen G.L.
        • Tang C.
        • et al.
        Radiotherapy-induced malfunction in contemporary cardiovascular implantable electronic devices: clinical incidence and predictors.
        JAMA Oncol. 2015; 1: 624-632
        • Bagur R.
        • Chamula M.
        • Brouillard E.
        • et al.
        Radiotherapy-induced cardiac implantable electronic device dysfunction in patients with cancer.
        Am. J. Cardiol. 2017; 119: 284-289